Table 3

New Risk Priority Numbers (RPNs)  and improvement programmes

Project namePotential causesFailure modeEffectsCorrective actionsNew OccurrenceNew
Detection
RPN changeIndicator
Traceability of the kitsTime between sampling and processing too long because of late delivery to the pharmacyDegraded specimenFalse negativeAmendment of instruction to the users, with request to register the date of sampling on the sampling tube2164 → 8Percentage FITs tested in laboratory within six calendar days of sampling (target value: 100%)
Time between sampling and processing too long because of late delivery from pharmacy to laboratoryControl of sampling date by the pharmacist
Registration of specimen’s code by the pharmacist
2164 → 8
Time between sampling and processing too long because of delay in the laboratoryAutomatic detection by the software of the difference between the date of delivery to a pharmacy and reading in the laboratory2164 → 8
Improving practices in the pharmacyWrong code because of manual entry by the pharmacistMixed-up kitsNegative test result never communicated to the userDouble-checking of the identity of the specimen’s owner by verbal request
Double-typing of the code by the pharmacist
Registration of the specimen’s code by the pharmacist
Self-certification of barcode-reading software by the pharmacy
2160 → 6Percentage of anonymous specimens (target value:<1.0%)
A person who has not done the test receives communication of a positive result2136 → 6Percentage of complaints about communication of a test result to a person who never took the test (target value: 0%)
A person who has not done the test receives communication of a negative result2136 → 6Percentage of complaints for the communication of a test result to a person who never took the test (target value: 0%)
Wrong code because of manual entry by the pharmacistAnonymous specimenNegative test result never communicated to the user2145 → 6Percentage of anonymous specimens (target value:<1.0%)
The pharmacist provides the kit without using the programme’s management softwareAnonymous specimenPositive test result never communicated to the user2148 → 8Percentage of anonymous specimens (target value:<1.0%)
Inadequate environmental temperature in the pharmacyDegraded specimenFalse negativeTraining for pharmacists on monitoring the storage temperature in the pharmacy1448 → 16Percentage FITs tested in laboratory within six calendar days of sampling (target value: 100%)
Lack of control of stocks by the pharmacistExpired sampling tubeUser dissatisfiedTraining for pharmacists on monitoring expiry dates1212 → 8Percentage of sampling tubes that have not expired (target value: 100%)
Improving information to usersInadequate environmental temperature at user’s homeDegraded specimenFalse negativeAmendment of instruction leaflet for users, including: correct sampling quantity, how to clean the tube, reporting of personal ID, collection modalities, delivery time after collection and preservation1448 → 16Leaflet amended
Improving laboratory practiceInadequate environmental temperature in the laboratoryDegraded specimenFalse negativeSharing and updating common procedures with the laboratory2236 → 16Regular meetings organised with the laboratory staff
  • FIT, faecal immunochemical test.