TY - JOUR T1 - Single-lead ECGs (AliveCor) are a feasible, cost-effective and safer alternative to 12-lead ECGs in community diagnosis and monitoring of atrial fibrillation JF - BMJ Open Quality JO - BMJ Open Qual DO - 10.1136/bmjoq-2020-001270 VL - 10 IS - 1 SP - e001270 AU - Jonathan James Hyett Bray AU - Elin Fflur Lloyd AU - Firdaus Adenwalla AU - Sarah Kelly AU - Kathie Wareham AU - Julian P J Halcox Y1 - 2021/03/01 UR - http://bmjopenquality.bmj.com/content/10/1/e001270.abstract N2 - Background Community management of atrial fibrillation (AF) often requires the use of electrocardiographic (ECG) investigation. Patients discharged following treatment of AF with fast ventricular response (fast AF) can require numerous ECGs to monitor rate and/or rhythm control. Single-lead ECGs have been proposed as a more convenient and relatively accurate alternative to 12-lead ECGs for rate/rhythm management and also diagnosis of AF. We aimed to examine the feasibility of using the AliveCor single-lead ECG monitor for diagnosis and monitoring of AF in the community setting.Methods During the course of 6 months, this evaluation of a clinical service improvement pathway used the AliveCor in management of patients requiring (1) follow-up ECGs for AF with previously documented rapid ventricular rate or (2) ECG confirmation of rhythm where AF was suspected. Twelve AliveCor devices provided to the acute community medical team were used to produce 30 s ECG rhythm strips (iECG) that were electronically sent to an overreading physician.Results Seventy-four patients (mean age 82 years) were managed on this pathway. (1) The AliveCor was successfully used to monitor the follow-up of 37 patients with fast AF, acquiring a combined total of 113 iECGs (median 1.5 ±3.75 per patient). None of these patients required a subsequent 12-lead ECG and this approach saved an estimate of up to £134.49 per patient. (2) Of 53 patients with abnormal pulses, the system helped identify 8 cases of new onset AF and 19 cases of previously known AF that had reverted from sinus back into AF.Conclusions We have demonstrated that the AliveCor system is a feasible, cost-effective, time-efficient and potentially safer alternative to serial 12-lead ECGs for community monitoring and diagnosis of AF. ER -