Scatter and transmission doses from several pediatric X-ray examinations in a nursery

  • Original Article
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

Background

While several studies have investigated the dose from scattered radiation from X-ray procedures in a pediatric nursery, they examined scatter from chest procedures only, or the types of examination were not specified.

Objective

The aim of this study was to collect scatter and transmission data from several types of X-ray examinations.

Materials and methods

Using a "newborn" anthropomorphic phantom and an ion chamber, a series of scatter and transmission dose measurements were performed using typical exposure factors for chest, chest and abdomen, skull, skeletal long bone and spine procedures. The phantom was inside a crib for all exposures.

Results

The maximum scatter dose measured at 1 m from the field center was about 0.05 µGy per exposure for lateral skulls. Transmission doses for lateral exams were around 0.1 µGy per exposure at 1 m from the isocenter.

Conclusions

The study demonstrated that scatter dose to other patients in a neonatal unit is not significant, assuming the distance between adjacent cribs is in the order of 1 m. Transmission doses are also low provided the beam is fully intercepted by the cassette. For an average workload the dose received by imaging technologists would be small.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1.
Fig. 2.

Similar content being viewed by others

References

  1. Iyer KR, Spitz L, Kiely E, et al (1995) Scattered radiation in a neonatal surgical unit. Eur J Pediatr Surg 5:286–287

    CAS  Google Scholar 

  2. Duetting T, Foerste B, Knoch T, et al (1999) Radiation exposure during chest X-ray examinations in a premature intensive care unit: phantom studies. Pediatr Radiol 29:158–162

    Article  PubMed  Google Scholar 

  3. Milkovic D, Ranogajec-Komor M, Ivanovi-Herceg Z, et al (2000) Doses in the vicinity of mobile X-ray equipment in a children's intensive care unit. Coll Antropol 24:235–240

    PubMed  Google Scholar 

  4. ICRP Publication 60: International Commission on Radiological Protection (1990). 1990 Recommendations of the International Commission on Radiological Protection

  5. Trout ED, Kelley JP (1972) Scattered radiation from a tissue-equivalent phantom for X rays from 50 to 300 kVp. Radiology 104:161–169

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to John W. Burrage.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Burrage, J.W., Rampant, P.L. & Beeson, B.P. Scatter and transmission doses from several pediatric X-ray examinations in a nursery. Pediatr Radiol 33, 704–708 (2003). https://doi.org/10.1007/s00247-003-0999-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-003-0999-1

Keywords

Navigation